Autonomous Monitoring Mode
Accessing Autonomous Mode
- Type mode from the primary menu and select option 2 and then configure the sensor
Autonomous Mode Operation
In autonomous mode, the sensor operates independently, following a pre-set measurement routine. Key details about this mode:
- Disabled Communications: All major communication functions to the sensor are disabled in this mode.
- Automatic Start: The sensor begins its pre-set scan routine immediately upon power-up.
- Audio Signals:
- On power-up, the sensor emits a single beep.
- 1–3 seconds later, it emits three additional beeps to confirm it has started operating.
- Streaming pH Data: When the sensor is connected to a terminal interface via an appropriate program, it streams live pH data.
Accessing Functions in Autonomous Mode
- Send the command
#700
through a terminal window. - This action pauses the current scan routine and allows for reconfiguration.
Returning to Autonomous Mode:
- Either repower the sensor, or send a reboot command
- After restarting, the sensor restarts its pre-set measurement routine.